2024-03-28T22:18:50Z
https://nagoya.repo.nii.ac.jp/oai
oai:nagoya.repo.nii.ac.jp:02005584
2023-05-09T00:18:00Z
1213:1620:1621:1681884086109
Measurement of breath hydrogen
呼気中水素測定の基礎的検討
劉, 鳳
LIU, Feng
近藤, 孝晴
KONDO, Takaharu
戸田, 安士
TODA, Yasushi
This investigation was performed as a part of the study for the incidence of milk intolerance among Japanese population. Breath hydrogen was analyzed with MicroLyzer (Quintron lnstruments). Successive dilutions of the standard gas and expired gas yielded a close correlation with a theoretical value. lntraanalysis variation was minimal and the value obtained with MicroLyzer showed a good correlation with the value obtained with a thermal conductivity gas chromatograph (Shimazu). All the subjects studied showed increased breath hydrogen levels as compatible with lactose deficiency but only 25% of the subjects had symptoms of lactose intolerance.
departmental bulletin paper
名古屋大学総合保健体育科学センター
The Research Center of Health, Physical Fitness and Sports, Nagoya University
1992-03-30
application/pdf
総合保健体育科学
1
15
33
37
Nagoya Journal of Health, Physical Fitness & Sports
0289-5412
https://nagoya.repo.nii.ac.jp/record/2005584/files/njhpfs_15_1_33.pdf
jpn