Radiation dose evaluation in tomosynthesis and C-arm cone-beam CT examinations with an anthropomorphic phantom

http://hdl.handle.net/2237/14184
http://hdl.handle.net/2237/14184
f1b3707d-447b-4ab0-a60e-8ce1d8e482f4
名前 / ファイル ライセンス アクション
Medical_Physics_37_8_4298.pdf Medical_Physics_37_8_4298.pdf (624.5 kB)
アイテムタイプ 学術雑誌論文 / Journal Article(1)
公開日 2010-09-22
タイトル
タイトル Radiation dose evaluation in tomosynthesis and C-arm cone-beam CT examinations with an anthropomorphic phantom
言語 en
著者 Koyama, Shuji

× Koyama, Shuji

WEKO 38910

en Koyama, Shuji

Search repository
Aoyama, Takahiko

× Aoyama, Takahiko

WEKO 38911

en Aoyama, Takahiko

Search repository
Oda, Nobuhiro

× Oda, Nobuhiro

WEKO 38912

en Oda, Nobuhiro

Search repository
Yamauchi-Kawaura, Chiyo

× Yamauchi-Kawaura, Chiyo

WEKO 38913

en Yamauchi-Kawaura, Chiyo

Search repository
アクセス権
アクセス権 open access
アクセス権URI http://purl.org/coar/access_right/c_abf2
権利
権利情報 Copyright (2010) American Institute of Physics. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics.
言語 en
キーワード
主題Scheme Other
主題 organ dose
キーワード
主題Scheme Other
主題 C-arm cone-beam CT
キーワード
主題Scheme Other
主題 tomosynthesis
キーワード
主題Scheme Other
主題 effective dose
キーワード
主題Scheme Other
主題 dosimetry
抄録
内容記述タイプ Abstract
内容記述 Purpose: The objective of this study was to evaluate organ dose and the effective dose to patients undergoing tomosynthesis (TS) and C-arm cone-beam computed tomography (CBCT) examinations and to compare the doses to those in multidetector CT (MDCT) scans. Methods: Patient doses were measured with small sized silicon-photodiode dosimeters, 48 in number, which were implanted at various tissue and organ positions within an anthropomorphic phantom. Output signals from photodiode dosimeters were read out on a personal computer, from which organ and effective doses were computed. The doses in head, chest, abdomen, and hip-joint TS, and in head and abdomen C-arm CBCT were evaluated for routine protocols on Shimadzu TS and C-arm CBCT systems, and the doses in MDCT with the same scan regions as in TS and CBCT were on Toshiba 64-detector-row CT scanners. Results: In TS examination of the head, chest, abdomen, and hip-joint, organ doses for organs within scan ranges were 1–4 mGy, and effective doses were 0.07 mSv for the head scan and around 1 mSv for other scans. In C-arm CBCT examinations of the head and abdomen, organ doses within scan range were 2–37 mGy, and effective doses were 1.2 mSv for the head scan and 4–5 mSv for abdominal scans. Effective doses in TS examinations were approximately a factor of 10 lower, while the doses in CBCT examinations were nearly the same level, compared to the doses in the corresponding MDCT examinations. Conclusions: TS examinations with low doses and excellent resolutions in coronal images compared to recent MDCT would widely be used in tomographic examinations of the chest, abdomen, pelvis, skeletal-joints, and knee instead of MDCT examinations with significantly high doses. Since patient dose in C-arm CBCT was nearly the same level as that in recent MDCT, the same consideration for high radiation dose would be required for the use of CBCT.
言語 en
出版者
出版者 American Institute of Physics
言語 en
言語
言語 eng
資源タイプ
資源タイプresource http://purl.org/coar/resource_type/c_6501
タイプ journal article
出版タイプ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
DOI
関連タイプ isVersionOf
識別子タイプ DOI
関連識別子 https://doi.org/10.1118/1.3465045
ISSN
収録物識別子タイプ PISSN
収録物識別子 0094-2405
書誌情報 en : Medical Physics

巻 37, 号 8, p. 4298-4306, 発行日 2009-08
フォーマット
値 application/pdf
著者版フラグ
値 publisher
URI
識別子 http://hdl.handle.net/2237/14184
識別子タイプ HDL
URI
識別子 http://dx.doi.org/10.1118/1.3465045
識別子タイプ DOI
戻る
0
views
See details
Views

Versions

Ver.1 2021-03-01 18:56:21.756952
Show All versions

Share

Share
tweet

Cite as

Other

print

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX
  • ZIP

コミュニティ

確認

確認

確認


Powered by WEKO3


Powered by WEKO3