@article{oai:nagoya.repo.nii.ac.jp:02001161, author = {Yamaji, Mayako and Ishiguchi, Tsuneo and Koyama, Shuji and Ikeda, Shuji and Kitagawa, Akira and Hagihara, Makiyo and Itoh, Yuji and Nakamura, Masaru and Ota, Toyohiro and Suzuki, Kojiro}, issue = {2}, journal = {Nagoya Journal of Medical Science}, month = {May}, note = {Distribution of radiation by C-arm cone-beam computed tomography (CBCT) in the angiographic suite and effectiveness of protection devices were assessed. CBCT image of a human phantom was obtained by a rotation of 220 degrees during 8 seconds of exposure. One hundred and twelve dosimeters were placed at different positions around the beam entry site, and color maps of dose distributions were drawn for horizontal and vertical planes. The measurements showed the highest radiation dose over 600 μGy by a single CBCT image acquisition at a distance of 60 cm from the beam entry site and a height of 90 cm from the floor. The color maps demonstrated the dose distribution to be more intense at the bilateral directions of the phantom. With the use of a ceiling-mounted transparent lead-acryl screen and a table-suspended lead curtain, the doses were reduced by 45–92 % at a direction of 210 degrees and a distance of 120 cm.}, pages = {277--286}, title = {Distribution of scatter radiation by C-arm cone-beam computed tomography in angiographic suite : measurement of doses and effectiveness of protection devices}, volume = {83}, year = {2021} }