Hemolytic anemia due to a felt strip used in the early stage of acute aortic dissection surgery |
|
言語 |
en |
| 著者 |
Kuwabara, Fumiaki
Yano, Daisuke
Mashiko, Yuji
Tajima, Kazuyoshi
Tokuda, Yoshiyuki
Hirate, Yuichi
|
| アクセス権 |
|
|
アクセス権 |
open access |
|
アクセス権URI |
http://purl.org/coar/access_right/c_abf2 |
| 権利 |
|
|
権利情報Resource |
http://creativecommons.org/licenses/by-nc-nd/4.0/ |
|
権利情報 |
Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International |
|
言語 |
en |
| キーワード |
|
|
主題Scheme |
Other |
|
主題 |
felt strip |
| キーワード |
|
|
主題Scheme |
Other |
|
主題 |
aortic dissection |
| キーワード |
|
|
主題Scheme |
Other |
|
主題 |
hemolytic anemia |
| 内容記述 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
Hemolytic anemia is a rare complication after aortic surgery. We herein report an early postoperative case of hemolytic anemia caused by an internal felt strip. A 57-year-old man underwent emergency partial aortic arch replacement for acute type A aortic dissection. The proximal stump was reinforced using internal and external polytetrafluoroethylene felt strips. The patient subsequently developed profound mechanical hemolytic anemia two weeks after the operation. Computed tomography did not reveal any narrowing of the anastomosis or kinking of the graft. However, transesophageal echocardiography confirmed that the internal felt strip had become inverted by the blood flow. Reoperation was performed to redo the proximal anastomosis, while also removing the internal felt strip. Hemolysis diminished soon after the reoperation. We encountered a case of acute aortic dissection that required reoperation because of hemolytic anemia caused by internal felt strip inversion. Further measures are required to prevent hemolysis with felt strips. |
|
言語 |
en |
| 出版者 |
|
|
出版者 |
Nagoya University Graduate School of Medicine, School of Medicine |
|
言語 |
en |
| 言語 |
|
|
言語 |
eng |
| 資源タイプ |
|
|
資源タイプresource |
http://purl.org/coar/resource_type/c_6501 |
|
タイプ |
departmental bulletin paper |
| 出版タイプ |
|
|
出版タイプ |
VoR |
|
出版タイプResource |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |
| ID登録 |
|
|
ID登録 |
10.18999/nagjms.87.3.590 |
|
ID登録タイプ |
JaLC |
| 関連情報 |
|
|
関連タイプ |
isVersionOf |
|
|
識別子タイプ |
URI |
|
|
関連識別子 |
https://www.med.nagoya-u.ac.jp/medlib/nagoya_j_med_sci/873.html |
| 収録物識別子 |
|
|
収録物識別子タイプ |
PISSN |
|
収録物識別子 |
0027-7622 |
| 収録物識別子 |
|
|
収録物識別子タイプ |
EISSN |
|
収録物識別子 |
2186-3326 |
| 書誌情報 |
en : Nagoya Journal of Medical Science
巻 87,
号 3,
p. 590-596,
発行日 2025-08
|