@article{oai:nagoya.repo.nii.ac.jp:00026268, author = {吉田, 澪代 and 加藤, 丈典 and 南, 雅代 and 榎並, 正樹 and Yoshida, Miyo and Kato, Takenori and Minami, Masayo and Enami, Masaki}, journal = {名古屋大学年代測定研究}, month = {Mar}, note = {The X-ray fluorescence (XRF) analyzer (Shimadzu XRF-1800) of the Division for Chronological Research, Institute for Space‒Earth Environmental Research, Nagoya University, was moved from the Furukawa Hall to the Research Institutes Building II on April 2016. To perform accurate and stable measurements of major elements with the XRF analyzer again after the moving, we examined the optimal analytical condition and remade calibration curves using 44 geochemical reference rock samples from the Geological Survey of Japan. The measurement result using the new calibration curves showed that the measured values were close to the reference values for all major elements except for ones with very low concentrations, indicating that accurate measurement of major elements has been enabled., 名古屋大学宇宙地球環境研究所年代測定研究部は2016年4月に古川記念館から研究所共同館IIに移転し、それに伴い蛍光X線分析装置(島津製作所製XRF-1800)も移設した。そのため、移転後のXRF装置の安定性、分析条件、検量線等について再検討する必要が生じた。本研究においては、岩石標準物質44試料を用いて主成分元素(Si、Ti、Al、Fe、Mn、Mg、Ca、Na、K、P)の検量線を作成し直し、この検量線を使用して岩石標準物質の分析を行った。その結果、いずれの元素においても分析値と推奨値の比の平均はすべて1に近く、また、ばらつきも実用上十分小さいことが示された。以上のことから、本XRF装置により、主成分元素の正確な測定が可能となった。}, pages = {66--71}, title = {蛍光X線分析法による岩石試料の定量化学分析 : (1)主成分元素\n}, volume = {1}, year = {2017} }