@article{oai:nagoya.repo.nii.ac.jp:00005350, author = {Toyokawa, H and Yoshizawa, M and Uritani, A and Mori, C and Takeda, N and Kudo, K}, issue = {3}, journal = {IEEE Transactions on nuclear science}, month = {Jul}, note = {A spherical neutron counter for spectroscopy and dosimetry has been developed. The counter consists of a spherical polyethylene moderator and three slender 3He position-sensitive proportional counters. Rough measurements of neutron fluence spectra and ambient dose equivalents, H^*(10), have been done in the moderated-neutron field of JAERI. Characteristics of the counter are examined and compared with those of a Bonner sphere neutron spectrometer.}, pages = {788--791}, title = {Performance of a Spherical Neutron Counter for Spectroscopy and Dosimetry}, volume = {44}, year = {1997} }