{"_buckets": {"deposit": "826d23b1-de0a-4b64-8979-8bdcfa7f48f2"}, "_deposit": {"id": "28222", "owners": [], "pid": {"revision_id": 0, "type": "depid", "value": "28222"}, "status": "published"}, "_oai": {"id": "oai:nagoya.repo.nii.ac.jp:00028222", "sets": ["501"]}, "author_link": ["91887", "91888", "91889", "91890", "91891", "91892", "91893"], "item_10_biblio_info_6": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2017-09", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "5", "bibliographicPageEnd": "278", "bibliographicPageStart": "271", "bibliographicVolumeNumber": "18", "bibliographic_titles": [{"bibliographic_title": "Journal of Applied Clinical Medical Physics", "bibliographic_titleLang": "en"}]}]}, "item_10_description_4": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "The American Association of Physicists in Medicine (AAPM) Working Group on TG‐51 published an Addendum to the AAPM\u0027s TG‐51 protocol (Addendum to TG‐51) in 2014, and the Japan Society of Medical Physics (JSMP) published a new dosimetry protocol JSMP 12 in 2012. In this study, we compared the absorbed dose to water determined at the reference depth for high‐energy photon beams following the recommendations given in AAPM TG‐51 and the Addendum to TG‐51, IAEA TRS‐398, and JSMP 12. This study was performed using measurements with flattened photon beams with nominal energies of 6 and 10 MV. Three widely used ionization chambers with different compositions, Exradin A12, PTW 30013, and IBA FC65‐P, were employed. Fully corrected charge readings obtained for the three chambers according to AAPM TG‐51 and the Addendum to TG‐51, which included the correction for the radiation beam profile (Prp), showed variations of 0.2% and 0.3% at 6 and 10 MV, respectively, from the readings corresponding to IAEA TRS‐398 and JSMP 12. The values for the beam quality conversion factor kQ obtained according to the three protocols agreed within 0.5%; the only exception was a 0.6% difference between the results obtained at 10 MV for Exradin A12 according to IAEA TRS‐398 and AAPM TG‐51 and the Addendum to TG‐51. Consequently, the values for the absorbed dose to water obtained for the three protocols agreed within 0.4%; the only exception was a 0.6% difference between the values obtained at 10 MV for PTW 30013 according to AAPM TG‐51 and the Addendum to TG‐51, and JSMP 12. While the difference in the absorbed dose to water determined by the three protocols depends on the kQ and Prp values, the absorbed dose to water obtained according to the three protocols agrees within the relative uncertainties for the three protocols.", "subitem_description_language": "en", "subitem_description_type": "Abstract"}]}, "item_10_publisher_32": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "American Association of Physicists in Medicine", "subitem_publisher_language": "en"}]}, "item_10_relation_11": {"attribute_name": "DOI", "attribute_value_mlt": [{"subitem_relation_type": "isVersionOf", "subitem_relation_type_id": {"subitem_relation_type_id_text": "https://doi.org/10.1002/acm2.12159", "subitem_relation_type_select": "DOI"}}]}, "item_10_rights_12": {"attribute_name": "権利", "attribute_value_mlt": [{"subitem_rights": "This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.", "subitem_rights_language": "en"}]}, "item_10_select_15": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_select_item": "publisher"}]}, "item_10_source_id_62": {"attribute_name": "ISSN(Online)", "attribute_value_mlt": [{"subitem_source_identifier": "1526-9914", "subitem_source_identifier_type": "EISSN"}]}, "item_1615787544753": {"attribute_name": "出版タイプ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_970fb48d4fbd8a85", "subitem_version_type": "VoR"}]}, "item_access_right": {"attribute_name": "アクセス権", "attribute_value_mlt": [{"subitem_access_right": "open access", "subitem_access_right_uri": "http://purl.org/coar/access_right/c_abf2"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Kinoshita, Naoki", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91887", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Oguchi, Hiroshi", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91888", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Nishimoto, Yasuhiro", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91889", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Adachi, Toshiki", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91890", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Shioura, Hiroki", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91891", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Kimura, Hirohiko", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91892", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Doi, Kunio", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "91893", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2019-06-12"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "Kinoshita_et_al-2017-Journal_of_Applied_Clinical_Medical_Physics.pdf", "filesize": [{"value": "103.3 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_note", "mimetype": "application/pdf", "size": 103300.0, "url": {"label": "Kinoshita_et_al-2017-Journal_of_Applied_Clinical_Medical_Physics", "objectType": "fulltext", "url": "https://nagoya.repo.nii.ac.jp/record/28222/files/Kinoshita_et_al-2017-Journal_of_Applied_Clinical_Medical_Physics.pdf"}, "version_id": "bde099fd-3e43-49df-9959-0256a97d0bb3"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "absorbed dose to water", "subitem_subject_scheme": "Other"}, {"subitem_subject": "Addendum to TG‐51 protocol", "subitem_subject_scheme": "Other"}, {"subitem_subject": "IAEA TRS‐398 protocol", "subitem_subject_scheme": "Other"}, {"subitem_subject": "JSMP 12 protocol", "subitem_subject_scheme": "Other"}, {"subitem_subject": "photon clinical reference dosimetry", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "journal article", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Comparison of AAPM Addendum to TG-51, IAEA TRS-398, and JSMP 12: Calibration of photon beams in water", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Comparison of AAPM Addendum to TG-51, IAEA TRS-398, and JSMP 12: Calibration of photon beams in water", "subitem_title_language": "en"}]}, "item_type_id": "10", "owner": "1", "path": ["501"], "permalink_uri": "http://hdl.handle.net/2237/00030418", "pubdate": {"attribute_name": "PubDate", "attribute_value": "2019-06-12"}, "publish_date": "2019-06-12", "publish_status": "0", "recid": "28222", "relation": {}, "relation_version_is_last": true, "title": ["Comparison of AAPM Addendum to TG-51, IAEA TRS-398, and JSMP 12: Calibration of photon beams in water"], "weko_shared_id": -1}
Comparison of AAPM Addendum to TG-51, IAEA TRS-398, and JSMP 12: Calibration of photon beams in water
http://hdl.handle.net/2237/00030418
http://hdl.handle.net/2237/00030418e936d967-f938-4dfe-8d94-5aadccf4ff9e
名前 / ファイル | ライセンス | アクション |
---|---|---|
Kinoshita_et_al-2017-Journal_of_Applied_Clinical_Medical_Physics (103.3 kB)
|
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-06-12 | |||||
タイトル | ||||||
タイトル | Comparison of AAPM Addendum to TG-51, IAEA TRS-398, and JSMP 12: Calibration of photon beams in water | |||||
言語 | en | |||||
著者 |
Kinoshita, Naoki
× Kinoshita, Naoki× Oguchi, Hiroshi× Nishimoto, Yasuhiro× Adachi, Toshiki× Shioura, Hiroki× Kimura, Hirohiko× Doi, Kunio |
|||||
アクセス権 | ||||||
アクセス権 | open access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_abf2 | |||||
権利 | ||||||
言語 | en | |||||
権利情報 | This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | absorbed dose to water | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | Addendum to TG‐51 protocol | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | IAEA TRS‐398 protocol | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | JSMP 12 protocol | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | photon clinical reference dosimetry | |||||
抄録 | ||||||
内容記述 | The American Association of Physicists in Medicine (AAPM) Working Group on TG‐51 published an Addendum to the AAPM's TG‐51 protocol (Addendum to TG‐51) in 2014, and the Japan Society of Medical Physics (JSMP) published a new dosimetry protocol JSMP 12 in 2012. In this study, we compared the absorbed dose to water determined at the reference depth for high‐energy photon beams following the recommendations given in AAPM TG‐51 and the Addendum to TG‐51, IAEA TRS‐398, and JSMP 12. This study was performed using measurements with flattened photon beams with nominal energies of 6 and 10 MV. Three widely used ionization chambers with different compositions, Exradin A12, PTW 30013, and IBA FC65‐P, were employed. Fully corrected charge readings obtained for the three chambers according to AAPM TG‐51 and the Addendum to TG‐51, which included the correction for the radiation beam profile (Prp), showed variations of 0.2% and 0.3% at 6 and 10 MV, respectively, from the readings corresponding to IAEA TRS‐398 and JSMP 12. The values for the beam quality conversion factor kQ obtained according to the three protocols agreed within 0.5%; the only exception was a 0.6% difference between the results obtained at 10 MV for Exradin A12 according to IAEA TRS‐398 and AAPM TG‐51 and the Addendum to TG‐51. Consequently, the values for the absorbed dose to water obtained for the three protocols agreed within 0.4%; the only exception was a 0.6% difference between the values obtained at 10 MV for PTW 30013 according to AAPM TG‐51 and the Addendum to TG‐51, and JSMP 12. While the difference in the absorbed dose to water determined by the three protocols depends on the kQ and Prp values, the absorbed dose to water obtained according to the three protocols agrees within the relative uncertainties for the three protocols. | |||||
言語 | en | |||||
内容記述タイプ | Abstract | |||||
出版者 | ||||||
言語 | en | |||||
出版者 | American Association of Physicists in Medicine | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプresource | http://purl.org/coar/resource_type/c_6501 | |||||
タイプ | journal article | |||||
出版タイプ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
DOI | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.1002/acm2.12159 | |||||
ISSN(Online) | ||||||
収録物識別子タイプ | EISSN | |||||
収録物識別子 | 1526-9914 | |||||
書誌情報 |
en : Journal of Applied Clinical Medical Physics 巻 18, 号 5, p. 271-278, 発行日 2017-09 |
|||||
著者版フラグ | ||||||
値 | publisher |