ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

{"_buckets": {"deposit": "a88ba091-1658-4c31-8845-c3cfca28e9ab"}, "_deposit": {"id": "29355", "owners": [], "pid": {"revision_id": 0, "type": "depid", "value": "29355"}, "status": "published"}, "_oai": {"id": "oai:nagoya.repo.nii.ac.jp:00029355", "sets": ["2478"]}, "item_9_biblio_info_6": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2020-02", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "1", "bibliographicPageEnd": "84", "bibliographicPageStart": "79", "bibliographicVolumeNumber": "82", "bibliographic_titles": [{"bibliographic_title": "Nagoya Journal of Medical Science"}]}]}, "item_9_description_4": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "In clinical practice, surgeons have stated that magnetic resonance imaging (MRI) can be performed in patients with titanium alloy implants. However, manufacturers and distributors of many implants may not comply with this common practice. As such, this study aimed to investigate manufacturers’ views on MRI use in patients fitted with their implants. The questionnaire survey was conducted between May and August 2018. Is your product compatible with MRI? ( ) Select from (1) to (3). In case of (1) or (2), up to ( ) Tesla. (1) MRI can be performed even at the sites of implanted fixators. (2) MRI can be performed at sites without implanted fixators. (3) MRI cannot be performed, or the manufacturer does not approve MRI use (cannot issue a certificate). The questionnaire forms were sent to 12 manufacturers, and the response rate was 100%. Manufacturers responded that they could not publicly allow MRI use in patients with their products. These findings do not conclude that MRI cannot be performed in such patients. This survey revealed that currently decisions regarding MRI use is left to the treating physicians. This situation poses a great problem for medical safety and imposes a substantial burden on physicians. As many problems remain in the field of orthopedic surgery, manufacturers of implants should proactively manage issues surrounding the usage of MRI.", "subitem_description_type": "Abstract"}]}, "item_9_identifier_registration": {"attribute_name": "ID登録", "attribute_value_mlt": [{"subitem_identifier_reg_text": "10.18999/nagjms.82.1.79", "subitem_identifier_reg_type": "JaLC"}]}, "item_9_publisher_32": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "Nagoya University Graduate School of Medicine, School of Medicine"}]}, "item_9_relation_43": {"attribute_name": "異版である", "attribute_value_mlt": [{"subitem_relation_type": "isVersionOf", "subitem_relation_type_id": {"subitem_relation_type_id_text": "http://www.med.nagoya-u.ac.jp/medlib/nagoya_j_med_sci/821.html ", "subitem_relation_type_select": "URI"}}]}, "item_9_select_15": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_select_item": "publisher"}]}, "item_9_source_id_61": {"attribute_name": "ISSN(Online)", "attribute_value_mlt": [{"subitem_source_identifier": "2186-3326", "subitem_source_identifier_type": "ISSN"}]}, "item_9_source_id_7": {"attribute_name": "ISSN(print)", "attribute_value_mlt": [{"subitem_source_identifier": "0027-7622", "subitem_source_identifier_type": "ISSN"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "Iwatsuki, Katsuyuki"}], "nameIdentifiers": [{"nameIdentifier": "96325", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Yoneda, Hidemasa"}], "nameIdentifiers": [{"nameIdentifier": "96326", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Onishi, Tetsuro"}], "nameIdentifiers": [{"nameIdentifier": "96327", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Ishii, Hisao"}], "nameIdentifiers": [{"nameIdentifier": "96328", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Kurimoto, Shigeru"}], "nameIdentifiers": [{"nameIdentifier": "96329", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Yamamoto, Michiro"}], "nameIdentifiers": [{"nameIdentifier": "96330", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Tatebe, Masahiro"}], "nameIdentifiers": [{"nameIdentifier": "96331", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "Hirata, Hitoshi"}], "nameIdentifiers": [{"nameIdentifier": "96332", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2020-02-26"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "10_Iwatsuki.pdf", "filesize": [{"value": "558.7 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 558700.0, "url": {"label": "10_Iwatsuki.pdf", "url": "https://nagoya.repo.nii.ac.jp/record/29355/files/10_Iwatsuki.pdf"}, "version_id": "603a8723-72a4-4d22-8678-8f1c348faf00"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "magnetic resonance imaging", "subitem_subject_scheme": "Other"}, {"subitem_subject": "implant", "subitem_subject_scheme": "Other"}, {"subitem_subject": "total arthroplasty", "subitem_subject_scheme": "Other"}, {"subitem_subject": "fracture", "subitem_subject_scheme": "Other"}, {"subitem_subject": "patient safety", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Compatibility of magnetic resonance imaging in patients with orthopedic implants: manufacturer questionnaires", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Compatibility of magnetic resonance imaging in patients with orthopedic implants: manufacturer questionnaires"}]}, "item_type_id": "9", "owner": "1", "path": ["2478"], "permalink_uri": "https://doi.org/10.18999/nagjms.82.1.79", "pubdate": {"attribute_name": "公開日", "attribute_value": "2020-02-26"}, "publish_date": "2020-02-26", "publish_status": "0", "recid": "29355", "relation": {}, "relation_version_is_last": true, "title": ["Compatibility of magnetic resonance imaging in patients with orthopedic implants: manufacturer questionnaires"], "weko_shared_id": 3}
  1. C100 医学部/医学系研究科
  2. C100b 紀要
  3. Nagoya journal of medical science
  4. 82(1)

Compatibility of magnetic resonance imaging in patients with orthopedic implants: manufacturer questionnaires

https://doi.org/10.18999/nagjms.82.1.79
https://doi.org/10.18999/nagjms.82.1.79
b6efad4f-d997-4c82-bdcd-e539a70f0a5e
名前 / ファイル ライセンス アクション
10_Iwatsuki.pdf 10_Iwatsuki.pdf (558.7 kB)
Item type 紀要論文 / Departmental Bulletin Paper(1)
公開日 2020-02-26
タイトル
タイトル Compatibility of magnetic resonance imaging in patients with orthopedic implants: manufacturer questionnaires
著者 Iwatsuki, Katsuyuki

× Iwatsuki, Katsuyuki

WEKO 96325

Iwatsuki, Katsuyuki

Search repository
Yoneda, Hidemasa

× Yoneda, Hidemasa

WEKO 96326

Yoneda, Hidemasa

Search repository
Onishi, Tetsuro

× Onishi, Tetsuro

WEKO 96327

Onishi, Tetsuro

Search repository
Ishii, Hisao

× Ishii, Hisao

WEKO 96328

Ishii, Hisao

Search repository
Kurimoto, Shigeru

× Kurimoto, Shigeru

WEKO 96329

Kurimoto, Shigeru

Search repository
Yamamoto, Michiro

× Yamamoto, Michiro

WEKO 96330

Yamamoto, Michiro

Search repository
Tatebe, Masahiro

× Tatebe, Masahiro

WEKO 96331

Tatebe, Masahiro

Search repository
Hirata, Hitoshi

× Hirata, Hitoshi

WEKO 96332

Hirata, Hitoshi

Search repository
キーワード
主題Scheme Other
主題 magnetic resonance imaging
キーワード
主題Scheme Other
主題 implant
キーワード
主題Scheme Other
主題 total arthroplasty
キーワード
主題Scheme Other
主題 fracture
キーワード
主題Scheme Other
主題 patient safety
抄録
内容記述 In clinical practice, surgeons have stated that magnetic resonance imaging (MRI) can be performed in patients with titanium alloy implants. However, manufacturers and distributors of many implants may not comply with this common practice. As such, this study aimed to investigate manufacturers’ views on MRI use in patients fitted with their implants. The questionnaire survey was conducted between May and August 2018. Is your product compatible with MRI? ( ) Select from (1) to (3). In case of (1) or (2), up to ( ) Tesla. (1) MRI can be performed even at the sites of implanted fixators. (2) MRI can be performed at sites without implanted fixators. (3) MRI cannot be performed, or the manufacturer does not approve MRI use (cannot issue a certificate). The questionnaire forms were sent to 12 manufacturers, and the response rate was 100%. Manufacturers responded that they could not publicly allow MRI use in patients with their products. These findings do not conclude that MRI cannot be performed in such patients. This survey revealed that currently decisions regarding MRI use is left to the treating physicians. This situation poses a great problem for medical safety and imposes a substantial burden on physicians. As many problems remain in the field of orthopedic surgery, manufacturers of implants should proactively manage issues surrounding the usage of MRI.
内容記述タイプ Abstract
出版者
出版者 Nagoya University Graduate School of Medicine, School of Medicine
言語
言語 eng
資源タイプ
資源 http://purl.org/coar/resource_type/c_6501
タイプ departmental bulletin paper
ID登録
ID登録 10.18999/nagjms.82.1.79
ID登録タイプ JaLC
関連情報
関連タイプ isVersionOf
識別子タイプ URI
関連識別子 http://www.med.nagoya-u.ac.jp/medlib/nagoya_j_med_sci/821.html
ISSN(print)
収録物識別子タイプ ISSN
収録物識別子 0027-7622
ISSN(Online)
収録物識別子タイプ ISSN
収録物識別子 2186-3326
書誌情報 Nagoya Journal of Medical Science

巻 82, 号 1, p. 79-84, 発行日 2020-02
著者版フラグ
値 publisher
戻る
0
views
See details
Views

Versions

Ver.1 2021-03-01 09:35:37.866457
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3