An intraoperative 3D image-based navigation error during cervical pedicle screw insertion

https://doi.org/10.18999/nagjms.82.4.799
https://doi.org/10.18999/nagjms.82.4.799
ad396aa6-238b-4994-a6fa-b9b68e0c1885
名前 / ファイル ライセンス アクション
20_Nakashima.pdf 20_Nakashima.pdf (3.2 MB)
license.icon
Item type 紀要論文 / Departmental Bulletin Paper(1)
公開日 2020-11-26
タイトル
タイトル An intraoperative 3D image-based navigation error during cervical pedicle screw insertion
言語 en
著者 Nakashima, Hiroaki

× Nakashima, Hiroaki

WEKO 102487

en Nakashima, Hiroaki

Search repository
Ishikawa, Yoshimoto

× Ishikawa, Yoshimoto

WEKO 102488

en Ishikawa, Yoshimoto

Search repository
Ando, Kei

× Ando, Kei

WEKO 102489

en Ando, Kei

Search repository
Kobayashi, Kazuyoshi

× Kobayashi, Kazuyoshi

WEKO 102490

en Kobayashi, Kazuyoshi

Search repository
Ishiguro, Naoki

× Ishiguro, Naoki

WEKO 102491

en Ishiguro, Naoki

Search repository
Imagama, Shiro

× Imagama, Shiro

WEKO 102492

en Imagama, Shiro

Search repository
アクセス権
アクセス権 open access
アクセス権URI http://purl.org/coar/access_right/c_abf2
権利
言語 en
権利情報Resource http://creativecommons.org/licenses/by-nc-nd/4.0/
権利情報 Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International
キーワード
主題Scheme Other
主題 cervical pedicle screw
キーワード
主題Scheme Other
主題 navigation
キーワード
主題Scheme Other
主題 O-armTM
キーワード
主題Scheme Other
主題 3D fluoroscopy
キーワード
主題Scheme Other
主題 navigation error
抄録
内容記述タイプ Abstract
内容記述 A 67-year-old man underwent posterior cervical decompression surgery for ossification of the posterior longitudinal ligament (OPLL) with fixation using cervical pedicle screws (CPSs) guided by intraoperative 3D image-based navigation. Intraoperatively, while creating the screw hole using the navigation probe, the virtual trajectory on the intraoperative navigation screen showed a 10-degree angle discrepancy in the axial plane depending on whether a probing force was or was not applied for making the hole. This was potentially caused by vertebra rotation and a bent probe. Consequently, the CPSs were placed more laterally than the ideal trajectory, which resulted in <2 mm lateral perforation to the foramen transversarium. There were no screw insertion-related perioperative complications. Based on this case, we conclude that navigation error during CPS insertion can occur even with intraoperative 3D image-based navigation. The risk of a bowed navigation probe caused by posterior cervical muscle and vertebra rotation should be considered, even with use of a navigation reference frame.
言語 en
出版者
出版者 Nagoya University Graduate School of Medicine, School of Medicine
言語 en
言語
言語 eng
資源タイプ
資源 http://purl.org/coar/resource_type/c_6501
タイプ departmental bulletin paper
出版タイプ
出版タイプ VoR
出版タイプResource http://purl.org/coar/version/c_970fb48d4fbd8a85
ID登録
ID登録 10.18999/nagjms.82.4.799
ID登録タイプ JaLC
関連情報
関連タイプ isVersionOf
識別子タイプ URI
関連識別子 http://www.med.nagoya-u.ac.jp/medlib/nagoya_j_med_sci/824.html
ISSN(print)
収録物識別子タイプ PISSN
収録物識別子 0027-7622
ISSN(Online)
収録物識別子タイプ EISSN
収録物識別子 2186-3326
書誌情報 en : Nagoya Journal of Medical Science

巻 82, 号 4, p. 799-805, 発行日 2020-11
著者版フラグ
値 publisher
戻る
0
views
See details
Views

Versions

Ver.1 2021-03-01 08:38:27.609462
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3