WEKO3
アイテム
Measurement and Evaluation of the Acoustic Noise of A 3 Tesla MR Scanner
https://doi.org/10.18999/nagjms.69.1-2.23
https://doi.org/10.18999/nagjms.69.1-2.2325dd4c15-51a6-4a90-af93-1fa8a9226d01
名前 / ファイル | ライセンス | アクション |
---|---|---|
p023-028.pdf (270.9 kB)
|
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2007-03-19 | |||||
タイトル | ||||||
タイトル | Measurement and Evaluation of the Acoustic Noise of A 3 Tesla MR Scanner | |||||
言語 | en | |||||
著者 |
Hattori, Yoko
× Hattori, Yoko× Fukatsu, Hiroshi× Ishigaki, Takeo |
|||||
アクセス権 | ||||||
アクセス権 | open access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_abf2 | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | Magnetic resonance (MR) | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | Acoustic noise | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | 3 Tesla MR | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | We measured the sound level and frequencies of the acoustic noise generated by a 3 Tesla (T) MR scanner, and investigated the subjective sound level for 30 healthy volunteers with either earplugs, headphones or both. The sound level of 3T was found to be higher than that of 1.5T in all sequences. The peak sound pressure level of 3T ranged from 125.7 dB for MR angiography to 130.7 dB for single shot EPI on the linear scale. The equivalent noise level was from 110.0 dB for FLAIR to 115.8 dB for T1-IR on the A-weighted scale, which exceeded 99 dB, the level regulated by the International Electrotechnical Commission (IEC). The study of the subjective sound level showed that the effect of noise reduction was not significantly different between earplugs and headphones. However, the use of both devices could reduce the subjective sound level significantly better than either one alone (P<0.01). Thus we propose wearing both devices for ear-protection during 3T examinations. | |||||
言語 | en | |||||
出版者 | ||||||
出版者 | Nagoya University School of Medicine | |||||
言語 | en | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
出版タイプ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
ID登録 | ||||||
ID登録 | 10.18999/nagjms.69.1-2.23 | |||||
ID登録タイプ | JaLC | |||||
ISSN(print) | ||||||
収録物識別子タイプ | PISSN | |||||
収録物識別子 | 0027-7622 | |||||
ISSN(Online) | ||||||
収録物識別子タイプ | EISSN | |||||
収録物識別子 | 2186-3326 | |||||
書誌情報 |
en : Nagoya Journal of Medical Science 巻 69, 号 1-2, p. 23-28, 発行日 2007-01 |
|||||
フォーマット | ||||||
値 | application/pdf | |||||
著者版フラグ | ||||||
値 | publisher | |||||
URI | ||||||
識別子 | http://hdl.handle.net/2237/7475 | |||||
識別子タイプ | HDL |